Re: Problem stsetch2 SNF 2009-09-13 19:01:53: red alarm : backside gas leak

Replaced the backside He cooling oring (lip seal) and cleaned the platten. Ran an O2 plasma for about 30 min. He leak rate was 20 mT/min.